Exercise 1 – Develop an acquisition method
Task 3. Find optimum fragmentor voltage for maximum response
20 Agilent 6400 Series Triple Quad LC/MS Familiarization Guide
5 Extract the remaining ion
chromatograms automatically.
Extract Defined Chromatograms
should be the default action for
Assign File Open Actions.
Open the remaining data files,
Sulfa_SIM80.d through
Sulfa_SIM220.d.
Close the Method Explorer.
a Select File Open Actions from the
General section in the Method
Explorer.
b Make sure that Actions to be run list
only contains Extract Defined
Chromatograms.
c Click File > Open Data File.
The system displays the Open Data
File dialog box.
d Select the data files to be opened,
Sulfa_SIM80.d through
Sulfa_SIM220.d.
e Mark the Run ‘File Open’ actions
from selected method check box.
(lower left corner)
The Qualitative Analysis Method
Editor lets you define actions to be
performed automatically upon
opening a data file(s).
Steps Detailed Instructions Comments