EasyManua.ls Logo

Agilent Technologies 6460 - Page 26

Agilent Technologies 6460
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Exercise 1 – Develop an acquisition method
Task 4. Determine product ion masses
26 Agilent 6400 Series Triple Quad LC/MS Familiarization Guide
n From the Method Explorer in the
Chromatogram section, click
Integrate (MS/MS).
o Select MS/MS as the Integrator
selection, if necessary.
These data files contain MS/MS
data, so you need to modify the
parameters in the Integrate
(MS/MS) section. If the data file
contained only MS data, you would
need to modify the parameters in
the Integrate (MS) section.
Figure 3 Integrate (MS/MS) > Integrator Tab
p Click the Peak Filters tab. Make sure
that the Limit (by height) to the
largest check box is marked and set to
the value
100 as shown below.
Figure 4 Integrate (MS/MS) > Peak Filters tab
Steps Detailed Instructions Comments

Related product manuals