Exercise 3 – Create a Triggered Dynamic MRM acquisition method
Task 3. Create a Triggered Dynamic MRM method from an existing database
Agilent 6400 Series Triple Quad LC/MS Familiarization Guide 57
3 Review the method in the Dynamic
MRM Viewer dialog box.
a Right-click the Scan segments table
and click Edit DMRM Method. The
Dynamic MRM Viewer dialog box is
opened.
b Type
200 for the Cycle time. This
value is shown in the Acquisition tab.
c Click between the Primaries only
button and the All transitions button
if the Dynamic MRM Statistics
information is not updating. Then,
click the All transitions button.
d Click Close.
The compounds in Optimizer were
overwritten by the compounds that
you updated in the Database
Browser program.
You can modify the Cycle time and
see how the Minimum Dwell Time
is changed. If the Minimum Dwell
Time is less than 5 ms, and
especially if it is less than 2 ms,
then signal-to-noise is poor.
A Dwell Time of 8 ms per transition
is fine.