Exercise 1 – Develop an acquisition method
Task 1. Enter acquisition parameters and acquire data
8 Agilent 6400 Series Triple Quad LC/MS Familiarization Guide
Figure 1 MassHunter Workstation – Data Acquisition window
Steps Detailed Instructions Comments
2 Enter MS parameters appropriate
for sulfa drug mix and save the
method as iiiMS2Scantest.m,
where iii are your initials.
See Table 3 on page 9.
a Click the QQQ tab in the Method
Editor window.
b Select MS2 Scan from the Scan Type
list in the Time Segments table.
c Enter the other MS parameters as
listed in Table 3. These parameters are
in either the Acquisition or the Source
tabs.
d Save the method as
iiiMS2Scantest.m, where iii are your
initials.