EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>6490

Agilent Technologies 6490 Guide Book

Agilent Technologies 6490
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
24 Agilent 6400 Series Triple Quad LC/MS Concepts Guide
2 Inner Workings – Triple Quadrupole MS versus Single Quadrupole MS
Innovative Enhancements in the 6490 Triple Quadrupole
Ions are generated using an electrospray ion source where the
analyte is simultaneously ionized and desolvated from the
liquid matrix. The iFunnel includes the application of Agilent
Jet Stream Technology (first introduced with the 6460) which
improves sensitivity via thermal gradient focusing and
enhanced desolvation.
The first innovative enhancement is the use of a short hexabore
capillary. It has 6 capillary inlets and samples up to 10X more
ion rich gas from the source. It captures the majority of the gas
from the source region. See Figure 10. The hexabore capillary
transmits a high gas/ion volume into the ion optic system.
Figure 9 The iFunnel Technology
Figure 10 Hexabore capillary

Other manuals for Agilent Technologies 6490

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6490 and is the answer not in the manual?

Agilent Technologies 6490 Specifications

General IconGeneral
BrandAgilent Technologies
Model6490
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals