Agilent Triple Quadrupole MS and Sensitivity 3
Linearity of the Agilent 6400 Series Triple Quadrupole MS
Agilent 6400 Series Triple Quad LC/MS Concepts Guide 45
Linearity of the Agilent 6400 Series Triple Quadrupole MS
Analyses of verapamil show the following linear dynamic range:
6490: six orders of magnitude (Figure 19)
6460, 6430, 6420: five orders of magnitude
Figure 19 Verapamil results - Calibration Curve 100 Attograms to 100 Picograms on-column,
Six orders of magnitude of linear dynamic range (log-log plot)
acquired with Agilent 6490 Triple Quadrupole with Agilent Jet Stream Technology