56 Agilent 6400 Series Triple Quad LC/MS Concepts Guide
3 Agilent Triple Quadrupole MS and Sensitivity 
Front-end ion optics 
Skimmer Agilent uses a small diameter skimmer orifice with very short 
distances from the capillary to the skimmer to the octopole. 
Consequently, more of the ions exiting the capillary are 
captured by the ion octopole guide, thereby keeping the ion 
beam very tightly focused. 
Octopole 1 Higher multipoles provide better ion transmission over a wider 
mass range. The depth of the potential well is steeper for the 
higher multipoles (at like voltages), especially those close to the 
rods resulting in the loss of fewer ions. This explains the 
predominant use of octopoles as ion guides in mass 
spectrometers where the main objective is to transmit rather 
than filter the ions. 
Lens 2 RF The phase of lens 2 is matched to that of the subsequent 
quadrupole, MS1 (Q1), resulting in a significantly increased 
sensitivity.
Quad mass 
filters
The quadrupoles consist of hyperbolic rods that optimize ion 
transmission and spectral resolution. There tends to be more 
ion loss with circular rods.
 Lens 2 RF (transmission of higher masses) Hyperbolic quadrupole and post-filter
Figure 25 Front-end optics