84
Maintaining the Instrument
This chapter describes the procedures for maintaining the instrument. Table 6
lists routine maintenance activities that should reduce unplanned repairs and
maintenance. The maintenance frequency appropriate for your instrument will
vary significantly depending on the sample matrix, solvents, temperatures, and
sample throughput.
Table 6. Recommended Maintenance Schedule
Maintenance Activity* Frequency*
Replace sample probe 3 month
Steam clean flow path 3 months
Check vial transport alignment and adjust if necessary 3 months
Check carousel alignment and adjust if necessary 6 months
PM Service and inlet maintenance 12 months
Flowpath / vent parts replacement 24 months
* These tasks and frequencies are based on experience. They are not based on life
time testing or benchmark testing.