Quick Reference 27
About Logs
Three logs are accessible from the keypad: the run log,
the maintenance log, and the system event log. To
access the logs, press [Logs] and toggle to the desired
log. The display will indicate the number of entries the
log contains. Scroll through the list.
Run log
The run log is cleared at the start of each new run.
During the current run, any deviations from the planned
method (including keypad entries) are listed in the run
log table. When the run log contains entries, the Run
Log LED lights.
Maintenance log
The maintenance log contains entries made by the
system when any of the user-defined component
counters reach a monitored limit. The log entry contains
a description of the counter, its current value, the
monitored limits, and which of its limits has been
reached. In addition, each user task related to the
counter is recorded in the log, including resetting,
enabling or disabling monitoring, and changing limits or
units (cycles or duration).
System event log
The system event log records significant events during
the GC’s operation. Some of the events also appear in
the run log if they are in effect during a run.