EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>8596E

Agilent Technologies 8596E User Manual

Agilent Technologies 8596E
673 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #140 background imageLoading...
Page #140 background image
Chapter 2 141
Making Adjustments
22. First LO Distribution Amplifier for Option 009 or 010
12.Check that the measuring receiver power level reads greater than
+12 dBm.
13.Disconnect the power sensor from LO OUTPUT, then reconnect the
50 termination to LO OUTPUT.
14.Disconnect the DMM leads from A10TP4 and A10TP2.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8596E and is the answer not in the manual?

Agilent Technologies 8596E Specifications

General IconGeneral
BrandAgilent Technologies
Model8596E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals