Service Guide 2-43
8719ET/20ET/22ET System Verification and Performance Tests
8719ES/20ES/22ES Performance Tests
RF Network Analyzers
3. Repeat step 2 (beginning with ) to enter the values for the remaining frequency
segments and power levels of your analyzer. Refer to Table 2-13 for the frequency
segment values.
4. When all of the frequency segments have been entered, press the following:
Test Setup
5. To set up the PORT 1 measurement (REFLECTION measurement on ET Models), press
the following:
For ES Models:
For ET Models:
Table 2-13 Frequency Segment Values
Segment Start
Frequency
Stop
Frequency
Number
of Points
All Models 1 50 MHz 50.000101 MHz 11
All Models 2 839.999899 MHz 840 MHz 11
All Models 3 0.84 GHz 8 GHz 201
8719ET/ES 4 8 GHz 13.51 GHz 101
8720ET/ES 4 8 GHz 20.05 GHz 151
8722ET/ES 4 8 GHz 20 GHz 151
8722ET/ES 5 20 GHz 40.05 GHz 201
ADD
DONE
LIST FREQ
ALL SEGS SWEEP
Meas
Trans: FWD S21 (B/R)
Format
LIN MAG
Avg
IF BW
10 x1
Scale Ref 2 x1
Meas
TRANSMISSN
Format
LIN MAG
Avg
IF BW
10 x1
Scale Ref 2 x1

Table of Contents

Other manuals for Agilent Technologies 8719ES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8719ES and is the answer not in the manual?

Agilent Technologies 8719ES Specifications

General IconGeneral
BrandAgilent Technologies
Model8719ES
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals