Chapter 4 Remote Interface Reference
Calibration Commands
88
CALibration:VOLTage:LEVel {MINimum | MIDdle|MAXimum}
This command can only be used after calibration is unsecured and the output
state is ON. It sets the power supply to a calibration point that is e ntered wit h
CAL:VOLT
command. During calibration, three points must be entered and the
low-end point (MIN) must b e selected and entered first.
CALibration:VOLTage:PROTection
This com mand cali brates the overvoltage protection circuit of the po wer
supply. It takes about 10 seconds to execute the comm and. The calibration
must be unsecured and the output be opened before calibrating the overvoltage
protection circuit. The power sup ply automati cally perform s the cali bration
and stores the new overvoltage constant in
nonvolatile
memory.
Notice that
voltage calibr ation pre ce des before sen ding this c omm a nd.