Chapter 8 Specifications
Performance Specifications
152
Transient Response Time
Less than 50 μsec for output to recover to within 15 mV following a change in output
current from full load to half load or vice versa
Command Processing Time
Average time for output voltage to begin to change after receipt of digital data when
the power supply is connected directly to the GPIB or RS-232 is less than 100 msec
OVP and OCP Accuracy, ±(% of output + offset)
OVP 0.5% + 0.5 V
OCP 0.5% + 0.5 A
Activation time : Average time for output to start to drop after OVP or OCP
condition occurs.
OVP <1.5 msec when the trip voltage is equal or greater than 3 V
<10 msec when the trip voltage is less than 3 V
OCP <10 msec