Chapter 3 Front-Panel Operation
System-Related Operations
43
3
Error Conditions
When the front-panel ERROR annunciator turns on, one or more command syntax or
hardware errors have been detected. A record of up to 20 errors can be stored in the
power supply’s error queue. See chapter 5 “Error Messages”, starting on
page 123 for
a complete listing of the errors.
Errors are retrieved in first-in-first-out (FIFO) order. The first error returned is the
first error that was stored. Errors are cleared as you read them over the remote
interface. When you have read all errors from the queue, the
ERROR annunciator
turns off. The power supply beeps once each time an error is generated.
If more than 20 errors have occurred when you operate the power supply over the
remote interface, the last error stored in the queue (the most recent error) is replaced
with -350, “Too many errors”. No additional errors are stored until you remove
errors from the queue. If no errors have occurred when you read the error queue,
the power supply responds with +0, “No error” over the remote interface or “NO
ERRORS” from the front panel.
The error queue is cleared when power has been off or after a *CLS (clear status)
command has been executed. The *RST (reset) command does not clear the error
queue.
Front-panel operation:
Press (Local) key to return to front-panel operation mode if the power supply
is in remote operation mode.
If the ERROR annunciator is on, press key to view the errors. Use the knob to
scroll through the error numbers. Press to view the text of the error message.
Press or key to increase or decrease the scrolling speed of the text on the
display. All errors are cleared when you exit the menu or let the display time out for
about 30 seconds.
Remote interface operation:
SYSTem:ERRor? Read and clear one error from the error queue
Errors have the following format (the error string may contain up to 80 characters).
-102, ‘‘Syntax error’’
3: ERR -102
Store
Local
Error
>
<
>

Table of Contents

Other manuals for Agilent Technologies E3634A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3634A and is the answer not in the manual?

Agilent Technologies E3634A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3634A
CategoryPower Supply
LanguageEnglish

Related product manuals