Chapter 7 Tutorial
Output Characteristics
137
7
Output Characteristics
An ideal constant-voltage power supply would have a zero output impedance at all
frequencies. Thus, as shown in Figure 7-3, the voltage would remain perfectly
constant in spite of any changes in output current demanded by the load.
Figure 7-3. Ideal Constant Voltage Figure 7-4. Ideal Constant Current
Power Supply Power Supply
The ideal constant-current power supply exhibits an infinite output impedance at all
frequencies. Thus as Figure 7-4 indicates, the ideal constant-current power supply
would accommodate a load resistance change by altering its output voltage by just
the amount necessary to maintain its output current at a constant value.
The output of the E3633A and E3634A power supplies can operate in either constant-
voltage (CV) mode or constant-current (CC) mode. Under certain fault conditions,
the power supply can not operate in either CV or CC mode and becomes unregulated.

Table of Contents

Other manuals for Agilent Technologies E3634A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3634A and is the answer not in the manual?

Agilent Technologies E3634A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3634A
CategoryPower Supply
LanguageEnglish

Related product manuals