Chapter 4 Remote Interface Reference
System-Related Commands
85
4
*TST?
This query performs a complete self-test of the power supply. Returns ‘‘0’’ if the self-
test passes or ‘‘1’’ or any non-zero value if it fails. If the self-test fails, an error message
is also generated with additional information on why the test failed.
*SAV { 1 | 2 | 3 }
This command stores the present state of the power supply to the specified location
in non-volatile memory. Three memory locations (numbered 1, 2 and 3) are available
to store operating states of the power supply. The state storage feature ‘‘remembers’’
the states or values of the following commands:
CURR, CURR:STEP, CURR:TRIG, CURR:PROT, CURR:PROT:STAT
DISP, OUTP, OUTP:REL, TRIG:DEL, TRIG:SOUR, VOLT,
VOLT:STEP, VOLT:TRIG, VOLT:PROT, VOLT:PROT:STAT, and
VOLT:RANG
To recall a stored state, you must use the same memory location used previously to
store the state.
*RCL { 1 | 2 | 3 }
This command recalls a previously stored state. To recall a stored state, you must use
the same memory location used previously to store the state.
Note DISP {OFF|ON} can be stored and recalled in remote interface mode only. Going
to local mode automatically sets the display state to ON.

Table of Contents

Other manuals for Agilent Technologies E3634A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3634A and is the answer not in the manual?

Agilent Technologies E3634A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3634A
CategoryPower Supply
LanguageEnglish

Related product manuals