Chapter 4
Front-Panel Key and SCPI Command Reference
Front-Panel Keys
146
Front-Panel Key and SCPI Command
Reference
4.2.5.5 RF Input Ranging
Sets the power input range of the PSA.
4.2.5.6 Input Atten
Sets the value of the input attenuation.
Mode: MRECEIVE
Key Path:
Input, RF Input Ranging
Remote Command: [:SENSe]:MRECeive:RAUTo OFF|ON|0|1
[:SENSe]:MRECeive:RAUTo?
Preset: OFF
State Saved: Saved in instrument state.
Dependencies/Couplings: This key is grayed out when current measurement is one of the four Audio
measurements.
Remote Command Notes: Global to the current mode.
Example: MREC:RAUT ON
MREC:RAUT?
Mode: MRECEIVE
Key Path:
Input, Input Attenuation
Remote Command: [:SENSe]:POWer:RF:ATTenuation <integer>
[:SENSe]:POWer:RF:ATTenuation?
Unit: dB
Preset: 0
State Saved: Saved in instrument state.
Min: 0 dB
Max: 70 dB
Remote Command Notes: Global to the current mode.
Example: POW:RF:ATT 30
POW:RF:ATT?

Table of Contents

Other manuals for Agilent Technologies E4445A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4445A and is the answer not in the manual?

Agilent Technologies E4445A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4445A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals