Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
180
Front-Panel Key and SCPI Command
Reference
4.4.2.2 Meas Setup
Accesses the Meas Setup menu.
4.4.2.2.1 Avg Number
Sets the number that will be averaged.
4.4.2.2.2 Avg Mode
Selects the type of termination control used for averaging. This determines the averaging action after the
specified number of average count is reached.
• Repeat - After reaching the average count, the averaging is reset and a new average is started
• Exponential - Each successive result after the average count is reached, is exponentially weighted and
combined with the existing average.
Mode: MRECEIVE
Key Path:
Meas Setup
Remote Command: [:SENSe]:RFPower:AVERage:COUNt <integer>
[:SENSe]:RFPower:AVERage:COUNt?
[:SENSe]:RFPower:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:RFPower:AVERage[:STATe]?
Preset: 4, ON
State Saved: Saved in instrument state.
Min: 1
Max: 1024
Example: :RFP:AVER:COUN 10
:RFP:AVER:COUN?
:RFP:AVER:STAT OFF
:RFP:AVER:STAT?
Mode: MRECEIVE
Key Path:
Meas Setup
Remote Command: [:SENSe]:RFPower:AVERage:TYPE REP|EXP
[:SENSe]:RFPower:AVERage:TYPE?
Preset: REP
State Saved: Saved in instrument state.
Range: Repeat | Exponential
Notes: None

Table of Contents

Other manuals for Agilent Technologies E4445A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4445A and is the answer not in the manual?

Agilent Technologies E4445A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4445A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals