Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
199
Front-Panel Key and SCPI Command
Reference
4.4.4.2.2.3 Ref Position
Allows you to set the display reference position to either Top, Ctr (center), or Bot (Bottom).
4.4.4.2.2.4 Scale Coupling
Allows you to toggle the scale coupling function between On and Off.
Notes: If the Scale Coupling is set to On, this value is automatically determined by the
measurement result. When you set this value manually, Scale Coupling
automatically changes to Off.
Example: :DISP:AMD:WIND:TRAC:Y:RLEV 0
:DISP:AMD:WIND:TRAC:Y:RLEV?
Mode: MRECEIVE
Key Path:
Amplitude Y Scale
Remote Command: :DISPlay:AMDepth[1]|2:WINDow[1]|2|3|4:TRACe:Y[:SCALe]:RPOSition
TOP|CENTer|BOTTom
:DISPlay:AMDepth[1]|2:WINDow[1]|2|3|4:TRACe:Y[:SCALe]:RPOSition
?
Preset: CENTer
State Saved: Saved in instrument state.
Range: Top | Ctr | Bottom
Example: :DISP:AMD:WIND:TRAC:Y:RPOS TOP
:DISP:AMD:WIND:TRAC:Y:RPOS?
Mode: MRECEIVE
Key Path:
Amplitude Y Scale
Remote Command: :DISPlay:AMDepth[1]|2:WINDow[1]|2|3|4:TRACe:Y[:SCALe]:COUPle
0|1|OFF|ON
:DISPlay:AMDepth[1]|2:WINDow[1]|2|3|4:TRACe:Y[:SCALe]:COUPle?
Preset: ON
State Saved: Saved in instrument state.
Range: On|Off
Dependencies/Couplings: See Restriction and Notes