Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
234
Front-Panel Key and SCPI Command
Reference
4.4.6.4.8 PM Dev Only
Allows you to select only the results of PM deviation. When it is NO, you can get the results of PM deviation,
modulation rate, distortion, and sinad at the same time. When it is YES, you can only get the results of PM
deviation. All others will return
− 999, which can help to increase the speed of the measurement.
4.4.6.4.9 Modulation Rate
Allows you to toggle the modulation rate result on and off. Turning the modulation rate off will help to improve
the measurement speed.
4.4.6.4.10 Modulation Distortion & SINAD
Allows you to toggle the modulation distortion and SINAD results on and off. Turning the modulation distortion
and SINAD results off will help to improve the measurement speed.
Key Path: Meas Setup, More
Remote Command: [:SENSe]:PMDeviation:ONLY YES|NO
[:SENSe]:PMDeviation:ONLY?
Default: NO
State Saved: Saved in instrument state.
Range YES|NO
Example: :PMD:ONLY NO
:PMD:ONLY?
Key Path:
Meas Setup, More
Remote Command: [:SENSe]:PMDeviation:MODRate ON|OFF
[:SENSe]:PMDeviation:MODRate?
Default: ON
State Saved: Saved in instrument state.
Range ON|OFF
Example: :PMD:MODR OFF
:PMD:MODR?
Key Path:
Meas Setup, More
Remote Command: [:SENSe]:PMDeviation:MODDist ON|OFF
[:SENSe]:PMDeviation:MODDist?