Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
251
Front-Panel Key and SCPI Command
Reference
The minimum bandwidth used is determined by the Low Pass Filter setting and is a value that is >
2xLowPassFilter. This mode usually works for most signals with a low modulation index. This mode will
increase measurement throughput because the analyzer can bypass the bandwidth measurement routines that
are performed at initialization for the Auto mode.
4.4.9.2.3 Average Number
Allows you to specify the number of measurement averages used when calculating the measurement result.
Key Path: Meas Setup
Remote Command: [:SENSe]:MODSinad:BANDwidth|BWIDth:TYPE MIN|AUTO|MAN
[:SENSe]:MODSinad:BANDwidth|BWIDth:TYPE?
Preset: AUTO
State Saved: Saved in instrument state.
Range: Auto|Man|Min
Notes: The IF BW type “Min” is meaningful only when the Modulation Mode is
“AM”. For FM, PM signal, the result will not be correct with this IF BW
Type.
Example: MODSinad:BAND:TYPE AUTO
MODSinad:BAND:TYPE?
Key Path:
Meas Setup
Remote Command: [:SENSe]:MODSinad:AVERage:COUNt <integer>
[:SENSe]:MODSinad:AVERage:COUNt?
[:SENSe]:MODSinad:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:MODSinad:AVERage[:STATe]?
Preset: 25, ON
State Saved: Saved in instrument state.
Min: 1
Max: 8192
Example: :MODSinad:AVERage:COUNt 25
:MODS:AVER:STAT OFF
:MODS:AVER:STAT?