4.4.6.4 Meas Setup
4.4.6.4.1 IF BW
Used specify the IF BW, if the IF BW Type is set to Man.
4.4.6.4.2 IF BW Type
Selects the type of using IF BW. It can be automatically set by measurement, or manually set by user. The
function of the tow types can reference to AM depth’s.
Key Path: Meas Setup
Remote Command: [:SENSe]:PMDeviation:BANDwidth|BWIDth <freq>
[:SENSe]:PMDeviation:BANDwidth|BWIDth?
Preset: 1MHz
State Saved: Saved in instrument state.
Range: 1 kHz to 10 MHz
Notes: When IF BW Type is set to “AUTO”, the measurement does not use this
value.
Example: :PMD:BAND 1MHz
:PMD:BAND?
Key Path:
Meas Setup
Remote Command: [:SENSe]:PMDeviation:BANDwidth|BWIDth:TYPE AUTO|MAN
[:SENSe]:PMDeviation:BANDwidth|BWIDth:TYPE?
Preset: AUTO
State Saved: Saved in instrument state.
Range: Auto | Man
Example: :PMD:BAND:TYPE MAN
:PMD:BAND:TYPE?