Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
231
Front-Panel Key and SCPI Command
Reference
4.4.6.4.3 Average Number
Sets the number that will be averaged. After the specified number (average counts) have been averaged, the
averaging mode (termination control) setting determines the averaging action.
4.4.6.4.4 Average Mode
Selects the type of termination control used for averaging. This determines the averaging action after the
specified number of frames (average count) is reached.
• Exponential - Each successive data acquisition after the average count is reached, is exponentially weighted
and combined with the existing average.
• Repeat - After reaching the average count, the averaging is reset and a new average is started.
Key Path: Meas Setup
Remote Command: [:SENSe]:PMDeviation:AVERage:COUNt <integer>
[:SENSe]:PMDeviation:AVERage:COUNt?
[:SENSe]:PMDeviation:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:PMDeviation:AVERage[:STATe]?
Preset: 25, ON
State Saved: Saved in instrument state.
Min: 1
Max: 8192
Example: :PMD:AVER:COUN 25
:PMD:AVER:COUN?
:PMD:AVER:STAT OFF
:PMD:AVER:STAT?
Key Path:
Meas Setup
Remote Command: [:SENSe]:PMDeviation:AVERage:TYPE REP|EXP
[:SENSe]:PMDeviation:AVERage:TYPE?
Preset: REP
State Saved: Saved in instrument state.
Range: Repeat | Exponential
Example: :PMD:AVER:TYPE EXP
:PMD:AVER:TYPE?