102 Chapter 3
Measurement Synchronization with LXI-B
Setting Up and Executing a List Sweep Measurement
15. Choose the MXAs
PTP domain
:LXI:CLOCk:PTP:DOMain 0 This command is sent
to the MXA.
The parameter value
should match the one
used in Task 3.
16. Choose the MXAs
LXI domain
:LXI:EVENt:DOMain 0 This command is sent
to the MXA.
The parameter value
should match the one
used in Task 4.
17. Turn off auto
alignments
CAL:AUTO:OFF This command is sent
to the MXA.
Auto alignments will
introduce timing
variability.
18. Put the MXA into
SA mode.
INST:SEL SA This command is sent
to the MXA.
19. Turn off continuous
measurements.
INIT:CONT OFF This command is sent
to the MXA.
20. Put the MXA into
the list sweep
measurement
CONF:LIST This command is sent
to the MXA.
21. Send the MXA a list
of frequencies.
LIST:FREQ 100 MHz, 200 MHz, 300
MHz, 400 MHz, 500 MHz
This command is sent
to the MXA.
22. Configure the
MXA’s list sweep
parameters.
LIST:ATT 10dB
LIST:BAND:RES:TYPE FLAT
LIST:BAND:RES 300 kHz
LIST:BAND:VID 3MHz
LIST:SWE:TIME 1e-6 s
LIST:TRIG:DEL 0
LIST:DET RMS
These commands are
sent to the MXA.
23. Configure the MXA
to trigger on an alarm.
:LIST:TRIGger:SOURce ALARM This command is sent
to the MXA.

Table of Contents

Other manuals for Agilent Technologies EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EXA N9010A and is the answer not in the manual?

Agilent Technologies EXA N9010A Specifications

General IconGeneral
BrandAgilent Technologies
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals