Chapter 3 93
Measurement Synchronization with LXI-B
Configuring LXI Triggers
Configuring an LXI State Trigger
Recalling a state in response to a peer to peer message is done in a
similar way, although only a SCPI interface is available.
Note that the Input Event Location parameter is also a part of the
instrumentâs state. This means that each saved state may point to new
states to be recalled on subsequent peer to peer messages. For example,
in âConfiguring an LXI State Triggerâ, Task 1 may include setting the
Location parameter for LAN1 to âc:\States\anotherState.stateâ. After
4. Monitor Response Follow âVerifying LXI Output
LAN Eventsâ on page 90.
This may be done
assuming âConfiguring
an LXI Trigger Using
SCPIâ on page 90 or
âConfiguring an LXI
Trigger Using the Front
Panelâ on page 92 are
complete.
Task Action Notes
1. Place the
instrument
in the state
to be recalled
Various SCPI commands may be used. This step is up to the userâs
discretion
2. Save the
state
:MMEM:STOR:STATe
âc:\States\myState.stateâ
The path must exist, and is up to
the userâs discretion.
3. Place the
instrument
in a new
state
Various SCPI commands may be used This step is up to the userâs
discretion
4. Set up the
state trigger
:LXI:EVENt:INPut:LAN:LOCation
âLAN0â,âc:\States\myState.stateâ
Parameter one may be any Event
Name reported by the query
:LXI:EVENt:INPut:LAN:LIST?
Parameter 2 must match the
path used in Task 2.
5. Enable the
state trigger
:LXI:EVENt:INPut:LAN:ENABled
âLAN0â,1