62 Chapter 1
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
bits that you want to enable.
Questionable Integrity Signal Positive Transition This command determines which bits in the
Questionable Integrity Signal Condition register will set the corresponding bit in the
Questionable Integrity Signal Event register when the condition register bit has a positive
transition (0 to 1). The variable <integer> is the sum of the decimal values of the bits that
you want to enable.
Questionable Integrity Uncalibrated Register
Questionable Integrity Uncalibrated Condition This query returns the decimal value of the
sum of the bits in the Questionable Integrity Uncalibrated Condition register.
NOTE The data in this register is continuously updated and reflects the
Mode All
Remote Command :STATus:QUEStionable:INTegrity:SIGNal:NTRansition
<integer>
:STATus:QUEStionable:INTegrity:SIGNal:NTRansition?
Example STAT:QUES:INT:SIGN:NTR 4 Burst found will be reported to the
Integrity Summary of the Status Questionable register.
Preset 0
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All
Remote Command :STATus:QUEStionable:INTegrity:SIGNal:PTRansition
<integer>
:STATus:QUEStionable:INTegrity:SIGNal:PTRansition?
Example STAT:QUES:INT:SIGN:PTR 4 Burst not found will be reported
to the Integrity Summary of the Status Questionable register.
Preset 32767
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command