Agilent 1200 Preparative Pump User Manual 123
Maintenance
7
Simple Repairs
2 Insert a hex-key in to hole in the multi assembly. Hold the hex-key firmly
and unscrew the locknut using the 7/16 inch wrench.
Figure 27 Changing the Filter Frit
3 Pull out the filter holder and filter.
4 Flush the filter port and filter holder using a wash bottle containing
isopropanol.
;^aiZg[g^i
=daZ
;^aiZg]daYZg