4 Using the Preparative Pump
Algae Growth in HPLC Systems
Algae growth may also be the possible source for failures of the ball valves
and other components in the flow path.
How to Prevent and/or Reduce the Algae Problem
Always use freshly prepared solvents, especially use demineralized water
which was filtered through about 0.2 µm filters.
Never leave mobile phase in the instrument for several days without flow.
Always discard “old” mobile phase.
Use the amber solvent bottle (part number 9301-1450) supplied with the
instrument for your aqueous mobile phase.
If possible add a few mg/l sodium azide or a few percent organic solvent to
the aqueous mobile phase.