164 Agilent 1200 Preparative Pump User Manual
9 Configuring the Preparative Pump
Main Power Supply Assembly
Main Power Supply Assembly
The main power supply comprises a closed assembly (no on-site repair
possibility).
The power supply provides all DC voltages used in the module except for the
voltages supplied by the lamp power supply to the deuterium and tungsten
lamps in the detectors. The line voltage can vary in a range from
100 240 volts AC ± 10% and needs no manual setting.
Figure 39 Main Power Supply (MPS) Blockdiagram
No accessible hardware fuse is needed because the main power supply is safe
against any short circuits or overload conditions on the output lines. When
overload conditions occur, the power supply turns off all output voltages.
Turning the line power off and on again resets the power supply to normal
operation if the cause of the overload condition has been removed.
+ 3 6 V
s u p p l y
+ 2 4 V
s u p p l y
+ 5 V
s w i t c h i n g
r e g u l a t o r
l i n e a r v o l t a g e
r e g u l a t o r s
+ 3 6 V ( n o t u s e d )
+ 2 4 V / 4 . 5 A
+ 5 V / 3
A
+ 1 5 V / . 3 A
- 1 5 V / . 3 A
l i n e
v o l t a g e
p r i m a r y
c i r c u i t s
M P S
s e n s e

Table of Contents

Other manuals for Agilent Technologies G1314B VWD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1314B VWD and is the answer not in the manual?

Agilent Technologies G1314B VWD Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1314B VWD
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals