Agilent 1200 Preparative Pump User Manual 27
Site Requirements and Specifications
2
Physical Specifications
Physical Specifications
Tabl e 4 Physical Specifications
Type Specification Comments
Weight 15.0 kg (34 lbs)
Dimensions
(width × depth × height)
200 x 345 x 440 mm (8 x 13.5 x 18
inches)
Line voltage 100 – 240 VAC, ± 10% Wide-ranging capability
Line frequency 50 or 60 Hz, ± 5%
Power consumption 250 VA / 210 W / 717 BTU Maximum
Ambient operating
temperature
4–40 °C (41–104 °F)
Ambient non-operating
temperature
-40–70 °C (-4–158 °F)
Humidity < 95%, at 25–40 °C (77–104 °F) Non-condensing
Operating Altitude Up to 2000 m (6500 ft)
Non-operating altitude Up to 4600 m (14950 ft) For storing the module
Safety standards: IEC, CSA,
UL
Installation Category II, Pollution
Degree 2
For indoor use only. Research
Use Only. Not for use in
Diagnostic Procedures.

Table of Contents

Other manuals for Agilent Technologies G1314B VWD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1314B VWD and is the answer not in the manual?

Agilent Technologies G1314B VWD Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1314B VWD
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals