90 Agilent 1200 Preparative Pump User Manual
6 Troubleshooting and Diagnostics
Error Messages
Wait for Volume Timed Out
The wait time limit until a given volume has been pumped, has been exceeded.
Wait for Pressure Timed Out
The limit, for waiting until a given pressure is reached, has been exceeded.
Composition Cluster Partner Lost
Communication to a composition cluster partner was interrupted.
In a gradient system, two (or more) isocratic preparative pumps can be linked
together in a cluster acting as one gradient pump. The error message is
generated, if the communication between the partners of a cluster is
interrupted.
Probable cause Suggested actions
1 Wrong limit set for the time limit. Set limit accordingly.
2 Flow setting too low. Set flow accordingly.
Probable cause Suggested actions
1 Wrong limit set for the time limit. Set limit accordingly.
2 Leak. Check for leaks with the help of the built in
pressure or leak test and correct them, where
necessary.
Probable cause Suggested actions
1 Composition cluster partner was turned OFF. Turn ON composition cluster partner.
2 CAN-cable to composition cluster partner
disconnected.
Reconnect or replace CAN-cable.
3 Defective CAN-bus. Replace PPM main board.

Table of Contents

Other manuals for Agilent Technologies G1314B VWD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1314B VWD and is the answer not in the manual?

Agilent Technologies G1314B VWD Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1314B VWD
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals