126 Agilent 1200 Preparative Pump User Manual
7 Maintenance
Simple Repairs
Removing and Disassembling the Pump Head Assembly
When When leaky or blocked
For exchanging the seals
For exchanging the plungers
Tools required Screwdriver Pozidrive#1
4-mm hexagonal key
10-mm wrench
Toothpick
Preparations Switch OFF preparative pump at power switch
Remove the front cover to have access to the pump mechanics
NOTE
Seals should always be replaced as pairs (2x front seals + 2x back seals)!
CAUTION
Damage of the pump drive
Starting the pump when the pump head is removed may damage the pump drive.
Never start the pump when the pump head is removed.