50 Agilent 1200 Preparative Pump User Manual
3 Installing the Pump
Get the System Ready for the First Analysis
5 When the pumping system has been turned OFF for a certain time (for
example, overnight) repeat Step 3 (step 4 on page 49).
Tabl e 10 Choice of Priming Solvents for Different Purposes
Activity Solvent Comments
After an installation
When switching between
reverse phase and normal
phase (both times)
Isopropanol Best solvent to flush air out of the
system
After an installation Ethanol or Methanol Alternative to Isopropanol (second
choice) if no Isopropanol is available
To clean the system when
using buffers
After a solvent change
Bidistilled water Best solvent to re-dissolve salts
After the installation of
normal phase solvents
Hexane + 5% Isopropanol Good wetting properties