EasyManuals Logo

Agilent Technologies gc/ms series User Manual

Agilent Technologies gc/ms series
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #139 background imageLoading...
Page #139 background image
137
Order online at www.agilent.com/chem/store
GC/MS Systems
The mass filter does not require periodic maintenance. It should not be removed from the
radiator or distributed in any way.
Never put the quadrupole in an ultrasonic cleaner.
Never change the physical orientation of the quadrupole mass filter.
The fused-quartz quadrupole is fragile and will break if dropped or handle roughly.
The material in the cusps of the quadrupole is very hygroscopic. If exposed to water,
the quadrupole must be dried very slowly to prevent damage.
Cleaning techniques appropriate for other manufacturers' instruments are not suitable
for Agilent MSDs – and may actually harm the mass filter.
To save time and effort, use only Agilent MSD mass filters, which do not require
periodic cleaning or maintenance.
In case of extreme contamination, contact a trained Agilent service representative to
perform the mass filter cleaning.
Quadrupole Mass Filter

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies gc/ms series and is the answer not in the manual?

Agilent Technologies gc/ms series Specifications

General IconGeneral
BrandAgilent Technologies
Modelgc/ms series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals