The mass filter does not require periodic maintenance. It should not be removed from the
radiator or distributed in any way.
Never put the quadrupole in an ultrasonic cleaner.
Never change the physical orientation of the quadrupole mass filter.
The fused-quartz quadrupole is fragile and will break if dropped or handle roughly.
The material in the cusps of the quadrupole is very hygroscopic. If exposed to water,
the quadrupole must be dried very slowly to prevent damage.
Cleaning techniques appropriate for other manufacturers' instruments are not suitable
for Agilent MSDs – and may actually harm the mass filter.
To save time and effort, use only Agilent MSD mass filters, which do not require
periodic cleaning or maintenance.
In case of extreme contamination, contact a trained Agilent service representative to
perform the mass filter cleaning.