4 Maintenance
Barcode errors
Seahorse XFe Analyzer Operating Manual 35
Cell plate barcode read failure
For any Cell Plate barcode read errors, Wave Controller displays a message and
two corrective actions. (See Figure 15.)
•Manual: Manually input the Cell Plate barcode info.
•Cancel Assay: Cancel the assay.
Manually enter cell plate barcode information as follows:
1 Click the Tray Widget. Cartridge Barcode Read Failure dialog box appears.
(See Figure 13 on page 33.)
2 To eject the Cell Plate, click Open Tray.
3 The Cell Plate barcode is located on the side of the plate. Write down the
barcode information.
4 Click Close Tray. The Cartridge Barcode Manual Entry window appears.
(See Figure 14 on page 34.)
5 Enter the Cell Plate barcode, and click Accept.
Figure 15. Cell Plate Barcode Read Failure dialog box