Setup and Calibration of the CP-8400/8410
CP-8400 AutoSampler & CP-8410 AutoInjector 15
To begin calibrating the CP-8400/8410, press the Calibrate softkey at the bottom
right corner of the Edit CP-8400 (or CP-8410) Setup screen. If you press Save
and Exit without calibrating, the existing calibration parameters will be retained.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
[ 2 ] Calibrate Injector 1 Position
[ 3 ] Calibrate Injector 2 Position
CP-8400 CALIBRATION SELECT
[ 1 ] Calibrate Vial 0 Position
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
[ 2 ] Calibrate Injector 1 Position
[ 3 ] Calibrate Injector 2 Position
CP-8410 CALIBRATION SELECT
[ 1 ] Calibrate Vial 1 Position
Press 1 on the CP-3800 keyboard to begin the “Calibrate Vial 0 Position”
procedure for the CP-8400. If you have a CP-8410, Press 1 on the CP-3800
keyboard to begin the “Calibrate Vial 1 Position” procedure for the CP-8410. This
procedure will allow the AutoSampler/AutoInjector to learn the tower and
carrousel/tray positions so that the syringe is positioned directly over the middle