Use
Switching on and Use of the TwisTorr 304 FS on
board
To switch on the TwisTorr 304 FS on board, it is sufficient to provide
the mains supply. The controller automatically recognizes the
presence of the interlock and start signals and starts the pump.
“Soft Start” mode is provided in order to switch on the pump after a
long period of standstill. To start the pump with “Soft Start” mode
active, this must be enabled via software.
For the “Soft Start” mode activation procedure, see the “Signal
Description” paragraph in the “Technical Information” chapter. The
flashing rate of the green LED on the TwisTorr 304 FS on board front
panel indicates system operating conditions:
with no flashing: the pump is rotating normally;
slow flashing (period of about 400 ms): the system is in ramp, or
in Stop, or in “Waiting for interlock” status;
fast flashing (period of about 200 ms): error condition.
Switching off the TwisTorr 304 FS on board
To switch off the TwisTorr 304 FS on board, it is sufficient to remove
the mains supply. The integrated controller immediately stops the
pump.
Emergency Stop
To stop the TwisTorr 304 FS on board immediately in an emergency
condition, remove the power cord.
138/178 TwisTorr 304 FS on board Controller User Manual / 87-901-028-01 (C)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies X3507-64002 and is the answer not in the manual?

Agilent Technologies X3507-64002 Specifications

General IconGeneral
BrandAgilent Technologies
ModelX3507-64002
CategoryWater Pump
LanguageEnglish

Related product manuals