Technical Information
Use in the Presence of Magnetic Fields
Use in the Presence of Magnetic Fields
Magnetic fields induce eddy currents in the rotor of a turbomolecular
pump that tend to prevent rotation. This causes increased electrical
power consumption by the motor, most of which is dissipated in the
rotor. In general, therefore, an increase in pump current can be
expected.
Orderable Parts
DESCRIPTION PART NUMBER
Mains cable NEMA Plug, 3m long 9699958
Mains cable European Plug, 3m long 9699957
Serial cable and T-Plus Software 9699883
Air cooling kit for TwisTorr 304 FS Controller
X3500-68010
Fan extension cable 9699949
Vent valve for TwisTorr 304 FS Controller 9699834
172/178 TwisTorr 304 FS on board Controller User Manual / 87-901-028-01 (C)