EasyManua.ls Logo

Ahlborn ALMEMO 710 - 10.4 Starting and stopping measuring operations

Ahlborn ALMEMO 710
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Starting and stopping measuring operations
10.4 Starting and stopping measuring operations
In addition to starting / stopping a cyclic measuring operation by means of the
Start / Stop keys or via the interface (see Manual 6.6) there are several
other methods, e.g. start time and end time, fixed measuring period, or limit
value actions.
Once a measuring operation has started (LED ´START´ lights up) no
programming is permitted. Similarly, once a measuring operation
has started, sensors must not be exchanged; a sensor connected in
mid-operation will not be recognized or evaluated.
The function ´Measuring start´ gives access to menu ´Settings´ > ´Data log-
ger 3/4´; here the ´Start mode´ can be set as manual or by start time. (see
10.4.1) To have the maximum / minimum / average values of all measuring
channels deleted each time a measuring operation is started, this can activated
in the next line (default). The ´Stop mode´ similarly can be set as manual or by
stop time or after a fixed measuring period. (see 10.4.2). To start / stop a
measuring operation as the result of a limit value infringement, ´Limit value
actions´ (s. 13.2.8) must be called up or the trigger variants listed in Chapter
13.5.2.
As soon as a measuring operation starts, the data logger display changes.
LED´s :
o START
o REC
o COM
Status Measuring operation in save-to-memory cycle | Scan mode
Measuring operation running since start time | Actual measuring duration since start
A measuring operation without memory activation will be as follows:
LED´s :
o START
o COM
Once-only output / saving of all measuring channels
A once-only, manual measuring channel scan to acquire the current measured
values from all active measuring channels (see Manual 6.5.1.1) can be per-
formed by pressing the key Manual . The results will be saved if the memory is
activated per scan cycle or per output cycle,
LEDs ´START´, ´REC´, ´COM´ light up briefly just once.
ALMEMO
®
710 33

Table of Contents

Other manuals for Ahlborn ALMEMO 710

Related product manuals