9 Calibration setup with more than 9 measuring locations
- 17 –
The numbering of the measuring locations in the
three-dimensional structure in the climatic
chamber is usually from left-front-top to right-
bottom-back. This arrangement must be
retained if the evaluation is to be carried out
later using the assistant for calibrating climatic
chambers in AMR WinControl.
Example input assignment on the ALMEMO
®
measuring device:
Measuring location Meas.
Measuring
Input
(Center) Psychrometer tw G0 M0
(Center) Psychrometer t, Uw, td, *R G0 M1
Air pressure sensor on the device p G0 M2
Meas. location 1 (Thermometer) t, Uw G0 M3
Meas. location 2 (Thermometer) t, Uw G0 M4
Meas. location 3 (Thermometer) t, Uw G0 M5
…
Meas. location 8 (Thermometer) t, Uw G0 M9
Meas. location 9 (Thermometer) MesR, t, Uw G1 M0
Meas. location 10 (Thermometer) t, Uw G1 M1
Meas. location 11 (Thermometer) t, Uw G1 M2
...
(Center) Thermometer (Radiation) t … …
(Center) Thermometer (Radiation) t … …