34 AS-D+—AS-P Series Instruction Manual
AirSep Corporation
5.2 PRE-INSTALLATION GUIDELINES
Before you install the oxygen generator and the oxygen receiver, if supplied, refer
to the Specications section in the Appendix of this instruction manual to determine
the applicable oor space, feed air, and the power requirements for your particular
model.
The oxygen generator may use feed air at specications outside
those shown in the Appendix of this instruction manual; however,
use of such feed air may require modication of the oxygen
generator at the AirSep Corporation factory to ensure the product
oxygen meets the specications. Consult your sales representative
to determine whether your oxygen generator requires modications
for your application.
Locate the oxygen generator in an area where the ambient air
temperature remains between 4°C (40°F) and 44°C (112°F) to
prevent damage not covered under the AirSep Corporation Product
Warranty.
To prevent re or electrical shock, locate the oxygen generator away
from rain or any other type of moisture.
Make sure the area that surrounds the oxygen generator is well
ventilated, and provide sufcient space around the unit (at least one
meter (three feet)) to allow for cool air ow as well as to allow safe
operation and maintenance.
AirSep oxygen generators are sold for use in commercial
applications only. Contact AirSep Corporation, or an authorized
AirSep representative, before you use this unit for any medical
application.
Do not disconnect protective earth:
Airsep oxygen generators are intended to be secured to the oor.
Failure to do so can result in the unit tipping over. An example of
securing to the oor would be bolting the unit down.