7.0 Troubleshooting
7.1 TECHNICAL SUPPORT
For assistance in troubleshooting or repairing the unit, or to order replacement parts,
contact the AirSep Commercial Products Service Department by telephone. In the
USA or Canada, call 1-800-320-0303. Outside the USA or Canada, call (716) 691-
0202. Send fax inquiries anytime to (716) 564-2717. Address written inquiries to:
AirSep Corporation
260 Creekside Drive
Buffalo, NY 14228-2075
USA
Attention: Commercial Products Service Department
E-mail: info@airsep.com
Visit www.airsepcpd.com to know about our complete range of standard Oxygen
Generators.
7.2 TROUBLESHOOTING CHART
Use the following chart as a guide to troubleshoot the oxygen generator.
Problem Probable Cause Solution
Oxygen generator does
not cycle.
No electrical power to
control cabinet.
Make sure unit is
plugged into wall outlet
and that wall outlet
receives power.
Circuit breaker tripped. Reset the breaker.
Generator is not
switched on.
Switch on the oxygen
generator.
Oxygen Generator is in
Standby mode.
The oxygen generator
will start running once
the pressure in the
oxygen receiver goes
below the low setpoint
for standby mode.