50 AS-D+—AS-P Series Instruction Manual
AirSep Corporation
Problem Probable Cause Solution
Oxygen generator
operates continuously.
Pressure of oxygen at
specied concentration
is lower than the
specied pressure.
Inadequate feed air
pressure.
Make sure feed air
valves remain fully
open and lines are not
obstructed.
Inadequate cycle
pressure.
Adjust or replace feed
air regulator.
Oxygen demand
exceeds oxygen
generator capacity.
Check oxygen usage or
check unit and oxygen
hoses and piping for
leaks.