Notes:
y To enter HEX values, the value must be preceded with a #h (i.e.
100 decimal = #h64).
y The device address and a read command can be concatenated on one
line if separated with a colon. Example:
C1;R? 100,1 (Set to read device #1, then read register at
100)
C2;R? 100,1 (Set to read device #2, then read register at
100)
y If using only one controller, it would be wise to set the device
address and then save it. Example:
C5
*SAV 0
Now the address will be remembered even after a power OFF
condition.
ALPHA OMEGA INSTRUMENTS CORP. Instruction Manual
Series 800/850 Thermoelectric Cooler Controller Page 89