EasyManuals Logo

Alpha Omega Instruments 800 Series User Manual

Default Icon
117 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #94 background imageLoading...
Page #94 background image
Notes:
y To enter HEX values, the value must be preceded with a #h (i.e.
100 decimal = #h64).
y The device address and a read command can be concatenated on one
line if separated with a colon. Example:
C1;R? 100,1 (Set to read device #1, then read register at
100)
C2;R? 100,1 (Set to read device #2, then read register at
100)
y If using only one controller, it would be wise to set the device
address and then save it. Example:
C5
*SAV 0
Now the address will be remembered even after a power OFF
condition.
ALPHA OMEGA INSTRUMENTS CORP. Instruction Manual
Series 800/850 Thermoelectric Cooler Controller Page 89

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Alpha Omega Instruments 800 Series and is the answer not in the manual?

Alpha Omega Instruments 800 Series Specifications

General IconGeneral
BrandAlpha Omega Instruments
Model800 Series
CategoryController
LanguageEnglish