EasyManua.ls Logo

Ambit PIB v1.5 - Page 14

Ambit PIB v1.5
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 1 • Introduction
10
bolus volume
11
, (4) lockout time, (5) volume to be infused and, if a PIB-IL v1.5 pump, (6) interval
limit.
The goal of the bolus or dose is to infuse the medication at an increased rate — mimicking an
injectionasmuchaspossible.Thepurposeoftheincreasedrateisgenerallytooodagreater
areawithanalgesicmedicationortoincreasetheconcentrationofthemedication.Theowrate
at which the ambIT
®
PIB v1.5 pumps infuse a bolus or dose is dependent on the energy state of
the batteries.
The bolus/dose infusion rate will be 210 ml/hr until the low battery alarm is activated if a
new, unused set of AA alkaline batteries are used at the start of the infusion. After the low
battery alarm is activated, the bolus/dose infusion rate may be reduced from the 210 ml/hr to
about 180 ml/hr. On a new, unused set of alkaline batteries, the low battery alarm will not be
reached until at least 500 ml have been infused. If one or both batteries have been previously
used, non-alkaline batteries are used, or the pump is repeatedly cycled off and on, the bolus/
dose infusion rate may drop to a minimum of 125 ml/hr prior to the low battery alarm being
activated.Thepurposeofallowingtheinfusionratetodropistomaximizebatterylife.The
pump will NOT inform the user that the bolus/dose infusion rate has dropped. No other
parameters are affected as the batteries are depleted.
CAUTION: If the medication being infused requires the bolus/dose infusion rate to be 210
ml/hr, only new alkaline batteries should be used and the batteries should be
changed immediately when the low battery alarm occurs.
CAUTION: Repeatedly cycling the pump off and on may cause the bolus/dose infusion
ratetodropbelow210ml/hr.Inextremecases,thebolus/doseinfusionrate
could drop as low as 125 ml/hr.
A new, unused set of alkaline batteries will infuse at 210 ml/hr. For a minimum of 500 ml, the
bolusanddoseowratewillbe210ml/hr.Oncethebatterieshavebeendepleted,theowrate
maydecrease.Thelowestowratethepumpwillreachis125ml/hr.
The different parameters for each mode interact with each other to create an infusion pattern.
ExamplesoftheinfusionpatternsforthedifferentmodesareprovidedgraphicallyinFigures
1-1through1-6thatfollow.Theowrates,volumes,lockouttimes,andintervalslistedare
examplesonly.
11 If the bolus volume is set to zero (0.0), then the lockout time is automatically bypassed and the pump will only deliver
doses (like PIb mode).

Table of Contents

Related product manuals