Section 3 • Programming Instructions
34
3.2.2.2 PIbModeProgrammingSteps
NOTE: If at any point during programming the user needs to move backward in the
display, refer to instructions found at 3.2.2.3.
ACTION
AUDIBLE
INDICATOR
VISUAL INDICATOR
1 Enterprogrammode:
From pause mode, press and hold
both the RUN/PAUSE button and
the BOLUS button until one beep
is heard.
One beep “PRO” is displayed for 2
seconds to indicate that the
pump is in program mode.
2 SelectPIb:
Press and release the RUN/
PAUSE button to scroll up or
press and release the BOLUS
button to scroll down through
program modes until “PIb”
appears in the display. To accept
the setting, press and hold the
RUN/PAUSE button until one
beep is heard.
One beep “PIb”ashesinthedisplay.
Mode is established.
3 Setdosevolume:
(A) Press and release the RUN/
PAUSE button to increase the
number or press and release the
BOLUS button to decrease the
number.
The letter “d,” the number
left of decimal point, and the
“ml”iconash.
(B) When the desired number
appears in the display, press and
hold the RUN/PAUSE button
until one beep is heard to accept
the setting and move onto the
nextdigit
One beep NOTE:Dosevolumeis
displayed in ml.
(C) Repeat steps 3(A) and 3(B)
to set the digit to the right of the
decimal point (00.0 to 00.9).
Letter “d” is displayed, and
number right of decimal
pointand“ml”ash.
Dose volume is established.
(D) If during programming, you
need to move backwards in the
display to a previous setting, press
and hold the BOLUS button until
one beep is heard.
Each time the BOLUS button
is pressed, the display moves
backward one screen to
enable review or correction.
NOTE:Thelimitsfordosevolumeare0.0to50.0ml.
ml
ml