Section 3 • Programming Instructions
35
ACTION
AUDIBLE
INDICATOR
VISUAL INDICATOR
4 Setdoseintervaltime:
(A) Press and release the RUN/
PAUSE button to increase the dose
interval time or press and release
the BOLUS button to decrease the
dose interval time.
“INT” is displayed for two
seconds to indicate that the
pump is in dose interval
mode.
“hr:min”andclockiconsand
thehourssectionash.
(B) When the desired dose interval
time appears in the display, press
and hold the RUN/PAUSE button
until one beep is heard to accept
the setting and move onto the
nextdigit.
NOTE:Doseintervaltime
is displayed in hours and
minutes(hh:mm).
(C) Repeat steps 4(A) and 4(B)
to set the digits in the minutes
section.
“hr:min”andclockiconsand
minutessectionash.
Dose interval time is
established.
(D) If during programming, you
need to move backwards in the
display to a previous setting, press
and hold the BOLUS button until
one beep is heard.
Each time the BOLUS button
is pressed, the display moves
backward one screen to
enable review or correction.
NOTE:Ifdosevolumeissetto“0,”thisstepwillbeskipped.
NOTE:Theminimumdoseintervaltimeisdependentonthedosevolumechosen.Themaximum
doseintervalisalways24:00(hh:mm).
5 Setvolumetobeinfused:
(A) Press and release the RUN/
PAUSE button to increase the
number or press and release the
BOLUS button to decrease the
number.
“ml” is displayed, and bag
icon and the hundreds digit
ash.
Volume to be infused is
displayed in ml.
(B) When the desired number
appears in the display, press and
hold the RUN/PAUSE button
until one beep is heard to accept
the setting and move onto the
nextdigit.
One beep “ml” is displayed, and bag
icon and the respective digits
ash.
ml
ml