Speciications | 2-3
Instrument Air Requirements
5 to 7 bar (70 to 100 psig)
Back Pressure Regulator Requirements (if purchased)
0 to 1 bar (0 to 15 psig) Maximum
Sample Inlet Pressure Requirement
138 kPa to 345 kPa (20 psig to 50 psig ) Maximum
(Pressure reduction provided by user.)
Exhaust Pressure
0 to 1 bar (0 to 15 psi) gauge. The minimum pressure drop across the 3050 must be 20 psi.
Minimum Differential Pressure
20 psi gauge (Inlet pressure must be greater than 20 psig over the outlet pressure.
Sample Flow Requirement
150 SCCM, 1 SLPM bypass low available for increased response speed.
Inlet Gas Temperature
0 °C to 60 °C (32 °F to 140 °F)
Optimal results are obtained when inlet gas temperature is maintained at 60 °C (140 °F).
Heat traced sample lines are recommended.
RTC Battery Speciication
Classiication: “Lithium Coin”
Chemical System: Lithium / Manganese Dioxide (Li/MnO2)
Nominal Voltage: 3.0 Volts
Typical Capacity: 235 mAh (to 2.0 volts)