EasyManua.ls Logo

Ametek 3050-SLR

Ametek 3050-SLR
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User Interface | 4-15
NOTE
nal low meter matches the external low meter. When calibration is complete press OK, to
return to the Monitor tab.
Test Alarms
The Test Alarms push buttons allow you to toggle the alarm contacts to an opened or closed
state. Use a multi-meter set to ohms to read resistance. Refer to igure 3-7 for location of
contacts on analyzer.
· Opened contacts should read ininity.
· Closed contacts should read zero.
Test mA Output
The Test mA Output push buttons allow you to test the analog outputs. Use a meter to test
output. Refer to igure 3-8 of for wiring.
To discontinue test mode, press another tab. Test mode will automatically time out after 10
minutes of inactivity.
When the analyzer switches from test mode to online, the analyzer resets itself.
Figure 4-11b.
Flow Adjust Screen.

Table of Contents

Related product manuals