C-4 5000 Multi Point Moisture Analyzer
Flags: Analog (Output 1-4)
Programmable Concentration Alert (1-2)
Display Lines (1,2, & 3)
0None
1 Point 1 ppm Held
2 Point 2 ppm Held
3 Point 3 ppm Held
4 Point 4 ppm Held
5 Roll through Points
Range Select-Out1
View Range Limits
Alter Range Limits
View Alarm Levels
Alter Alarm Levels
Calib
Alarm
Analog
Range
Test
Config
Start Auto Cal
Flow Moisture Gen (Manual Cal)
Test Display
Tst Analog Out 1&3
Tst Analog Out 2&4
Test Analog Inpts
Test Alarm Relays
Test Valve Relays
Test Watchdog
Test Serial Port
Test RAM
Information contained on this sheet subject to revision.
PN 500084001 Rev B / December 2005
Alarm_1_Hi
Alarm_2_Hi
Alarm_3_Hi
Alarm_4_Hi
Alarm_1_Low
Alarm_2_Low
Alarm_3_Low
Alarm_4_Low
Max_Cal_Err
Max_Span Max_CD
Output_1_Low Offset_1_Low
Output_1_Hi Offset_1_Hi
Output_2 Offset_2
Output_3 Offset_3
Output_4 Offset_4
Display Line ONLY Flags
12 Display Selected Streams
14 Alarm/Valve Signal Display
15 Timers
16 Output Range (Channel 1)
Flags1: Enter sum of flag values for
functions to be enabled:
1 Enable Auto-Range for Analog Output 1
2 Use Range Alert for "data valid" signal
4 Use Concentration Alert for "data valid" signal
8 Make Alarm Contacts Normally Open
16 NO Span Adjustment During Auto-Cal Cycle
32 Make Outputs Track During Auto-Cal
64 Display Negative Moisture Value as Zero
128 Use Non-Verbose Display Mode on Line 4
(Alarm Info Only)
Flags2: Enter sum of flag values for
functions to be enabled:
1 Special Applications Only
2 Special Applications Only
4 Use Enhanced Noise Smoothing
(Boxcar Averaging)
8 Special Applications Only
5000 Controller
with Multi-Point
6 Current Moisture (ppm)
7 Cell Frequency (Hz)
8 Delta Frequency (Hz)
View Configuration
Select Streams
Set Time of Day
Set Date
Set Auto Cal Time
System Tests Menu
Configure Menu
Configure Level
Security
Service Level
Security
PolyCoef Label
Constants Flags2
LVL2 Conf Timers
Coef_D
Default Memory
Constants Display
Analog
Passwords Timing
Moist_Gen Serial
Flags1 Noise
See Reverse Side for Configure Menu Maps
Refer to flag lists at right for input values for menu items in bold type.